Calibration date and how often
The camera's timing clock (its time base) drifts slowly as it ages, so a timer-enabled EtherLynx camera should be recertified regularly. FinishLynx shows the date of the last calibration, so you can check any camera, including a loan unit.
How to fix it
Check the last calibration date
- Boot the camera in FinishLynx
- Open Camera Settings and click the Setup tab
- The Information line shows the camera's serial number and its calibration date
If FinishLynx shows a Timer tab
- When a camera's date is past four years, or unknown (n/a), FinishLynx adds a Timer tab to Camera Settings as a reminder. The camera keeps timing
- The hidden setting Hardware\Ethernet\CheckCalib turns the check off, or on for every boot. Turning it off hides the reminder but doesn't change whether the camera is due. See the calibration check card on the FinishLynx Other Settings page
- The reminder, and how the check works, are shown in FinishLynx 11.22: Camera Calibration & Certification
How often
- Send the camera for service and recertification at least every four years
- Some governing bodies and championship meets require it more often. Check their rules if you time sanctioned events
To get it done
Request it through the Camera Calibration page. It explains what the recertification covers and how to send the camera in. If the camera is going in for repair anyway, ask for certification on the same RMA.