Capture button not working
For the capture button to work, its driver must be installed so Windows gives it a COM port, that COM port must be selected in FinishLynx, and the button must be plugged in before FinishLynx opens. On a new computer, the missing piece is usually the driver.
How to fix it
Which capture button do you have?
- Older model: a 9-pin serial plug, connected to the computer through the ATEN USB Serial Adapter (UC232A). It needs the ATEN USB Serial Adapter Driver
- Newer model: USB all the way to the button, no adapter. It needs the USB Capture Button Driver
- Either way the button plugs into the computer, not the camera. A cable with a 3-pin XLR on one end and a 15-pin plug on the other is the camera's DB15-XLR start cable, not a capture button cord
Install the driver
- Connect the computer to the internet, plug the button in and restart the computer. Windows may install the driver on its own
- If it doesn't, download and install the driver for your model from the Drivers & Utilities section of the Software page: the USB Capture Button Driver for the newer button, or the ATEN USB Serial Adapter Driver for the older button and its adapter
Find the COM port in Device Manager
- With the button plugged in, open Device Manager and expand Ports (COM & LPT)
- Note the COM number shown for the button or adapter (for example, COM3)
- If it shows a warning symbol, or there is no COM port, plug it into a different USB port and reinstall the driver
Select the port in FinishLynx
- Plug the capture button in before you open FinishLynx. FinishLynx reads the COM ports when it opens
- Click File | Options | General tab
- In Capture Button, select the COM port of the button, then click OK
- If the COM port isn't in the list, close and reopen FinishLynx with the button plugged in
- Still missing? See My USB adapter is installed, but the COM port isn't listed and the COM port card on the Other Settings page
Test it
- Open or create an event. The Capture icon shows capture is activated
- Press and hold the capture button. FinishLynx captures while the button is held and stops when you release it
- If FinishLynx sees the port but pressing the button does nothing, open Camera Settings | Capture tab and make sure Manual is checked
- If the camera now captures all the time, see Camera captures nonstop
Windows 11 replaced the driver
On some Windows 11 computers, Windows overrides the capture button's USB serial driver or fails to install it. The button works for a few events and then stops, and Device Manager shows a warning symbol on the adapter.
- Reinstall the driver for your model, then check Device Manager for a COM number with no warning symbol and select it again in FinishLynx
- If it keeps happening with the older adapter model, that adapter may not be compatible with Windows 11. The options are:
- The current USB capture button cable (no adapter, Windows 11 compatible), from Lynx
- A USB-to-serial adapter that is compatible with Windows 11
- The Automatic Capture plug-in, so you don't need the button. Plug-ins aren't available in FinishLynx LITE (see LITE vs full). If Automatic is grayed out after installing it, see Option grayed out or missing
- Use the button below to get the cable or the plug-in
ARM (Snapdragon) laptop
- Check the processor: Settings | System | About | System type. If it says ARM-based processor, the standard capture button driver may not load, and Device Manager shows the USB serial device with no driver loaded
- Install the ARM64 version of the FTDI driver from FTDI's driver page (FTDI makes the USB chip in the newer button). This has fixed it. Then select the COM port in File | Options | General tab
- If you can't get the driver loaded, use the button below before your meet
Works, then stops mid-meet
If the button works when the event loads but stops after a few races, and restarting brings it back, Windows is putting the USB port to sleep to save power.
- Open Device Manager | Universal Serial Bus controllers and double-click each USB Root Hub
- On the Power Management tab, uncheck "Allow the computer to turn off this device to save power" and click OK. Repeat for every USB Root Hub listed
- Turn off USB selective suspend: Control Panel | Power Options | Change plan settings | Change advanced power settings | USB settings | USB selective suspend setting | Disabled
- On a Windows 11 computer, also check Device Manager for a warning symbol on the button (see above)
Meet-day workaround, keyboard capture
- In File | Options | General tab, set Capture Button to Keyboard and click OK
- Open the event and click the Capture icon. The Keyboard Capture dialog opens
- Click the dialog so it has focus, then tap the Space bar once to start capturing and once to stop. Don't hold the key down
- While keyboard capture is on, the keyboard can't be used for anything else. For the "Spacebar toggles capture" message, see Spacebar toggles capture
The Manual Capture page of the manual covers both the button and keyboard capture. If the button still doesn't work after these checks, or never worked, use the button below.